Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Ventri Discovery NM 530c Recalled by GE Healthcare, LLC Due to GE Healthcare has recently become aware of an...

Date: January 9, 2013
Company: GE Healthcare, LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact GE Healthcare, LLC directly.

Affected Products

Ventri Discovery NM 530c, Ventri with Premium Table Discovery NM 530c, models H3000ZW & H3000SA. The system is used to perform nuclear imaging procedures.

Quantity: 556

Why Was This Recalled?

GE Healthcare has recently become aware of an incident which caused a patient injury during an unload following an exam on a Ventri scanner. During a patient unload following a scan, the patients fingers may be pinched between the moving cradle edge and the stationary base of the table while the table is moving outward automatically.

Where Was This Sold?

This product was distributed to 40 states: AL, AZ, AR, CA, CO, CT, DE, FL, IL, IN, IA, KY, LA, ME, MD, MA, MI, MS, MO, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WV, WI, DC

Affected (40 states)Not affected

About GE Healthcare, LLC

GE Healthcare, LLC has 474 total recalls tracked by RecallDetector.

Related Recalls