Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Touchscreen display kit. Sold under the following product names: TOUCHSCREEN Recalled by GE Healthcare, LLC Due to Upgraded Aisys anesthesia devices and Aisys CS2 anesthesia...

Date: September 27, 2018
Company: GE Healthcare, LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact GE Healthcare, LLC directly.

Affected Products

Touchscreen display kit. Sold under the following product names: TOUCHSCREEN DISPLAY KIT 11.X ENGLISH FOR USA, TOUCHSCREEN DISPLAY KIT 11.X ENGLISH, TOUCHSCREEN DISPLAY KIT 11.X GERMAN, TOUCHSCREEN DISPLAY KIT 11.X DUTCH, TOUCHSCREEN DISPLAY KIT 11.X FRENCH, TOUCHSCREEN DISPLAY KIT 11.X DANISH, TOUCHSCREEN DISPLAY KIT 11.X ESTONIAN, TOUCHSCREEN DISPLAY KIT 11.X ITALIAN, TOUCHSCREEN DISPLAY KIT 11.X NORWEGIAN. Product Usage: The GE Datex-Ohmeda Aisys Anesthesia System is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). The device is intended for volume or pressure control ventilation. The Aisys is not suitable for use in a MRI environment.

Quantity: 4,207 units total

Why Was This Recalled?

Upgraded Aisys anesthesia devices and Aisys CS2 anesthesia devices with PSVPro Spontaneous Breathing Modes do not provide the PSVPro ventilation mode feature of transitioning to SIMV PCV backup ventilation mode. PSVPro Spontaneous Breathing Modes do not provide the feature of transitioning to SIMV PCV backup ventilation mode for a patient who stops spontaneous breathing if the PSVPro ventilation mode is in use prior to the clinician starting the Cycling Procedure.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About GE Healthcare, LLC

GE Healthcare, LLC has 474 total recalls tracked by RecallDetector.

Related Recalls