Replacement leadwires with grabber ends: REF/Catalog Number Description 2106388-005 Replacement Recalled by GE Healthcare, LLC Due to ECG trunk cables and leadwires may reduce the...

Date: May 10, 2019
Company: GE Healthcare, LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact GE Healthcare, LLC directly.

Affected Products

Replacement leadwires with grabber ends: REF/Catalog Number Description 2106388-005 Replacement ECG Leadwire, grabber, BRN V, AHA, 74 cm/ 29 in 2106388-101 Replacement ECG Leadwire, grabber, LL , AHA, 130 cm/ 51 in 2106388-103 Replacement ECG Leadwire, grabber, GRN RL, AHA, 130 cm/ 2106388-107 Replacement ECG Leadwire, grabber, BLK N, IEC, 130 cm/ 51 in 2106392-004 Replacement ECG Leadwire, grabber, ORG V, AHA, 74 cm/ 29 in The reusable ECG leadwire sets enable ECG monitoring with GE equipment requiring rectangular 11-pin ECG trunk cable. Use leadwire sets with ECG trunk cable or telemetry transmitter.

Quantity: 58,045 units (31,108 US; 26937 OUS) in total

Why Was This Recalled?

ECG trunk cables and leadwires may reduce the amount of energy reaching the patient during defibrillation, potentially limiting success of defibrillating the patients rhythm.