Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

NM/CT 860 Nuclear Medicine / CT Scanners Recalled by GE Healthcare, LLC Due to GE Healthcare has become aware of a potential...

Date: February 7, 2021
Company: GE Healthcare, LLC
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact GE Healthcare, LLC directly.

Affected Products

NM/CT 860 Nuclear Medicine / CT Scanners

Quantity: 242 in total (17 units US)

Why Was This Recalled?

GE Healthcare has become aware of a potential issue on 800 family NM/CT Scanners that could cause a shift in CT radiation exposure range of up to 5cm compared to the intended radiation exposure range of the planned scan under specific workflows. The issue occurs only on a hybrid whole-body continuous F3 protocol with Zoom < 1 where the scan range is set on the Smart Console. In some cases, this may also necessitate a re-scan of the patient, which would expose the patient to additional X-ray radiation.

Where Was This Sold?

This product was distributed nationwide across all 50 states.