Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
NM/CT 850 Nuclear Medicine / CT Scanners Recalled by GE Healthcare, LLC Due to GE Healthcare has become aware of a potential...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact GE Healthcare, LLC directly.
Affected Products
NM/CT 850 Nuclear Medicine / CT Scanners
Quantity: 242 in total (17 units US)
Why Was This Recalled?
GE Healthcare has become aware of a potential issue on 800 family NM/CT Scanners that could cause a shift in CT radiation exposure range of up to 5cm compared to the intended radiation exposure range of the planned scan under specific workflows. The issue occurs only on a hybrid whole-body continuous F3 protocol with Zoom < 1 where the scan range is set on the Smart Console. In some cases, this may also necessitate a re-scan of the patient, which would expose the patient to additional X-ray radiation.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About GE Healthcare, LLC
GE Healthcare, LLC has 474 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026