Class III — Low Risk

Low risk — use of or exposure to this product is not likely to cause adverse health consequences.

Leadwire sets with snap ends: REF/Catalog Number Description 2106381-001 ECG Recalled by GE Healthcare, LLC Due to ECG trunk cables and leadwires may reduce the...

Date: May 10, 2019
Company: GE Healthcare, LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact GE Healthcare, LLC directly.

Affected Products

Leadwire sets with snap ends: REF/Catalog Number Description 2106381-001 ECG Leadwire set, 5-lead, snap, AHA, 74 cm/ 29 in 2106381-002 ECG Leadwire set, 5-lead, snap, AHA, 130 cm/ 51 in 2106381-003 ECG Leadwire set, 5-lead, snap, IEC, 74 cm/ 29 in 2106381-004 ECG Leadwire set, 5-lead, snap, IEC, 130 cm/ 51 in 2106381-005 ECG Leadwire set, 5-lead, snap, AHA, mix 74 cm/ 29 in, 130 cm/ 51 in 2106383-001 ECG Leadwire set, 5-lead, grouped, snap, AHA, 74 cm/ 29 in 2106383-002 ECG Leadwire set, 5-lead, grouped, snap, AHA, 130 cm/ 51 in 2106383-003 ECG Leadwire set, 5-lead, grouped, snap, IEC, 74 cm/ 29 in 2106383-004 ECG Leadwire set, 5-lead, grouped, snap, IEC, 130 cm/ 51 in 2106383-005 ECG Leadwire set, 5-lead, grouped, snap, AHA, mix 74 cm/ 29 in, 130 cm/ 51 in 2106383-006 ECG Leadwire set, 5-lead, grouped, snap, IEC, mix 74 cm/ 29 in, 130 cm/ 51 in 2106385-001 ECG Leadwire set, 3-lead, snap, AHA, 74 cm/ 29 in 2106385-002 ECG LEADWIRE SET, 3-LD GROUPED, SNAP, AHA, 130 CM/ 51 IN 2106385-003 ECG LEADWIRE SET, 3-LD GROUPED, SNAP, IEC, 74 CM/ 29 IN 2106385-004 ECG LEADWIRE SET, 3-LD GROUPED, SNAP, IEC, 130 CM/ 51 IN 2106394-003 ECG Leadwire set, 5-lead V2-6, snap, AHA, 74 cm/ 29 in 2106395-001 ECG Leadwire set, 6-lead, grouped, snap, AHA, 74 cm/ 29 in 2106396-001 ECG Leadwire set, 6-lead, grouped, snap, IEC, 74 cm/ 29 in 2106396-002 ECG Leadwire set, 6-lead, grouped, snap, IEC, 130 cm/ 51 in The reusable ECG leadwire sets enable ECG monitoring with GE equipment requiring rectangular 11-pin ECG trunk cable. Use leadwire sets with ECG trunk cable or telemetry transmitter.

Quantity: 58,045 units (31,108 US; 26937 OUS) in total

Why Was This Recalled?

ECG trunk cables and leadwires may reduce the amount of energy reaching the patient during defibrillation, potentially limiting success of defibrillating the patients rhythm.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About GE Healthcare, LLC

GE Healthcare, LLC has 474 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report