Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

GE Healthcare Recalled by GE Healthcare, LLC Due to GE Healthcare has recently become aware of a...

Date: June 8, 2012
Company: GE Healthcare, LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact GE Healthcare, LLC directly.

Affected Products

GE Healthcare, Optima MR360/Brivo MR355 (HDsv), GE 0.35T Signa Ovation HD MR System, GE 0.2T Signa Profile HD MR System. Indicated for use as a diagnostic imaging device.

Quantity: 864 (57 US; 802 OUS)

Why Was This Recalled?

GE Healthcare has recently become aware of a Reformat Image Flip Issue associated with GE MR Products. The axial and coronal reformatted images from a sagittal 3D series (such as with applications of FSPGR-3D T1, BRAVO T1, or CUBE T1) of over 126 slices may display a left-to-right image flip . In addition, the acquired sagittal images may be annotated with left-right location annotation revers

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About GE Healthcare, LLC

GE Healthcare, LLC has 474 total recalls tracked by RecallDetector.