Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

GE Healthcare Recalled by GE Healthcare, LLC Due to When the E-NMT-01 module is used in conjunction...

Date: April 9, 2014
Company: GE Healthcare, LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact GE Healthcare, LLC directly.

Affected Products

GE Healthcare, NeuroMuscular Transmission Module, E-NMT-01 Product Usage: The Datex-Ohmeda S/5TM E-NMT module is indicated for monitoring the relaxation of the patient and regional block stimulation for nerve location.

Quantity: 1,794 (471 units US, 1,323 units OUS)

Why Was This Recalled?

When the E-NMT-01 module is used in conjunction with the ElectroSensor, the Neuromuscular Transmission (NMT) values may indicate a deeper level of muscle relaxation than the actual level of muscle relaxation. In the clinical situation visual movements of the hand are seen after TOF (Train of Four) stimulation, but the patient monitor shows no counts, or counts are not corresponding to the actual a

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About GE Healthcare, LLC

GE Healthcare, LLC has 474 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report