Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

GE Healthcare Recalled by GE Healthcare, LLC Due to When performing a secondary capture generated from GSI...

Date: March 25, 2011
Company: GE Healthcare, LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact GE Healthcare, LLC directly.

Affected Products

GE Healthcare, Model Number/ Description: 5342114 GSI Viewer 1.10 on AW VolumeShare 4 (AW 4.5); 5342114-2 GSI Viewer 1.10 on AW VolumeShare 4 (AW 4.5); 5342114-3 GSI Viewer 1.10 on AW VolumeShare 4 (AW 4.5); 5342114-4 GSI Viewer 1.11 on AW VolumeShare 4 (AW 4.5); 5401757 GSI Viewer 1.20 on AW VolumeShare 3 (AW4.4). Computed Tomography X-Ray system

Quantity: 283 (59 US, 224 OUS)

Why Was This Recalled?

When performing a secondary capture generated from GSI Viewer, the header information is correct on the images. However, when changing the ROI (region of interest) and zooming in, the DICOM header does not change to reflect the new image, thus the line measurements may be inaccurate.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About GE Healthcare, LLC

GE Healthcare, LLC has 474 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report