Low risk — use of or exposure to this product is not likely to cause adverse health consequences.
ECG leadwire sets: REF/Catalog Number Description 2106389-001 ECG Leadwire set Recalled by GE Healthcare, LLC Due to ECG trunk cables and leadwires may reduce the...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact GE Healthcare, LLC directly.
Affected Products
ECG leadwire sets: REF/Catalog Number Description 2106389-001 ECG Leadwire set, 5-lead, grouped, grabber, AHA, 74 cm/ 29 in 2106389-002 ECG Leadwire set, 5-lead, grouped, grabber, AHA, 130 cm/ 51 in 2106389-003 ECG Leadwire set, 5-lead, grouped, grabber, IEC, 74 cm/ 29 in 2106389-004 ECG Leadwire set, 5-lead, grouped, grabber, IEC, 130 cm/ 51 in 2106389-005 ECG Leadwire set, 5-lead, grouped, grabber, AHA, mix 74 cm/ 29 in, 130 cm/ 51 in 2106389-006 ECG Leadwire set, 5-lead, grouped, grabber, IEC, mix 74 cm/ 29 in, 130 cm/ 51 in 2106390-001 ECG Leadwire set, 3-lead, grabber, AHA, 74 cm/ 29 in 2106390-002 ECG Leadwire set, 3-lead, grabber, AHA, 130 cm/ 51 in 2106390-003 ECG Leadwire set, 3-lead, grabber, IEC, 74 cm/ 29 in 2106390-004 ECG Leadwire set, 3-lead, grabber, IEC, 130 cm/ 51 in 2106391-001 ECG Leadwire set, 5-lead, grabber, AHA, 74 cm/ 29 in 2106391-002 ECG Leadwire set, 5-lead, grabber, AHA, 130 cm/ 51 in 2106391-003 ECG Leadwire set, 5-lead, grabber, IEC, 74 cm/ 29 in 2106391-004 ECG Leadwire set, 5-lead, grabber, IEC, 130 cm/ 51 in 2106391-005 ECG Leadwire set, 5-lead, grabber, AHA, mix 74 cm/ 29 in, 130 cm/ 51 in 2106391-006 ECG Leadwire set, 5-lead, grabber, IEC, mix 74 cm/ 29 in, 130 cm/ 51 in 2106393-001 ECG Leadwire set, 5-lead V2-6, grabber, AHA, 130 cm/ 51 in 2106393-002 ECG Leadwire set, 5-lead C2-6, grabber, IEC, 130 cm/ 51 in 2106393-003 ECG Leadwire set, 5-lead V2-6, grabber, AHA, 74 cm/ 29 in 2106393-004 ECG Leadwire set, 5-lead C2-6, grabber, IEC, 74 cm/ 29 in 2106397-001 ECG Leadwire set, 6-lead, grouped, grabber, AHA, 74 cm/ 29 in 2106397-002 ECG Leadwire set, 6-lead, grouped, grabber, AHA, 130 cm/ 51 in 2106397-003 ECG Leadwire set, 6-lead, grouped, grabber, AHA, mix 74 cm/ 29 in, 130 cm/ 51 in 2106398-001 ECG Leadwire set, 6-lead, grouped, grabber, IEC, 74 cm/ 29 in 2106398-002 ECG Leadwire set, 6-lead, grouped, grabber, IEC, 130 cm/ 51 in The reusable ECG leadwire sets enable ECG monitoring with GE equipment requiring rectangular 11-pin ECG trunk cable. Use leadwire sets with ECG trunk cable or telemetry transmitter.
Quantity: 58,045 units (31,108 US; 26937 OUS) in total
Why Was This Recalled?
ECG trunk cables and leadwires may reduce the amount of energy reaching the patient during defibrillation, potentially limiting success of defibrillating the patients rhythm.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About GE Healthcare, LLC
GE Healthcare, LLC has 474 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report