Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Discovery MI Digital Ready Recalled by GE Healthcare, LLC Due to DIGITAL ReadyCT/PET systems recently installed with a GE...

Date: May 24, 2018
Company: GE Healthcare, LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact GE Healthcare, LLC directly.

Affected Products

Discovery MI Digital Ready

Quantity: 42 units total

Why Was This Recalled?

DIGITAL ReadyCT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About GE Healthcare, LLC

GE Healthcare, LLC has 474 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report