Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Discovery 670 DR (Model H3100BT) Recalled by GE Healthcare, LLC Due to GE Healthcare has become aware that rotor bearing...

Date: December 10, 2020
Company: GE Healthcare, LLC
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact GE Healthcare, LLC directly.

Affected Products

Discovery 670 DR (Model H3100BT), NM/CT 870 DR (Model 3100Q and H3100AS), Discovery NM670 Pro (Model 3100XB), Discovery 670 ES (Model H3100XF) - Product Usage: intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such NM System: General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality.

Quantity: 52 untits

Why Was This Recalled?

GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.

Where Was This Sold?

Worldwide

About GE Healthcare, LLC

GE Healthcare, LLC has 474 total recalls tracked by RecallDetector.