Kenex Radiation Shield installed with GE Healthcare Interventional Recalled by GE Healthcare Due to GE Healthcare has recently become aware of an...

Date: March 9, 2015
Company: GE Healthcare
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact GE Healthcare directly.

Affected Products

Kenex Radiation Shield installed with GE Healthcare Interventional, Surgery, CT and X-Ray Imaging systems.

Quantity: 797 (3 US, 794 OUS).

Why Was This Recalled?

GE Healthcare has recently become aware of an improper installation of a ceiling suspended pendent supporting a radiation shield and/or surgical lamp. The pendent system was not installed per specified installation requirements. There has been a reported incident of a radiation shield falling from the ceiling causing a minor injury.