Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

GE Healthcare Recalled by GE Healthcare Due to GE Healthcare has recently become aware of a...

Date: December 8, 2014
Company: GE Healthcare
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact GE Healthcare directly.

Affected Products

GE Healthcare, Proteus XR/a with wall stand model number 600-0301.

Quantity: 3096 (2486 US, 610 OUS).

Why Was This Recalled?

GE Healthcare has recently become aware of a potential safety issue involving the breakage of the two steel cables which support the bucky device in the wall stand of Proteus XR/a X-ray imaging systems. No injuries reported.

Where Was This Sold?

This product was distributed to 1 state: DC

Affected (1 state)Not affected

About GE Healthcare

GE Healthcare has 88 total recalls tracked by RecallDetector.