Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

GE Healthcare Centricty PACS Recalled by GE Healthcare Due to GE HealthCare has become aware of an issue...

Date: August 5, 2024
Company: GE Healthcare
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact GE Healthcare directly.

Affected Products

GE Healthcare Centricty PACS, V2, V3, V4, V6, V7; Radiological imaging processing system.

Quantity: 2138 units

Why Was This Recalled?

GE HealthCare has become aware of an issue where image storage is not possible when Centricity PACS reaches the maximum number of unique image identifiers (Ckeys). If this occurs, it can result in acquisition of a partial study.

Where Was This Sold?

Worldwide distribution.