Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

GE Healthcare Automatic Mobile X-Ray (AMX) Series: Optima XR200 amx Recalled by GE Healthcare Due to An issue was identified with the mAs accuracy...

Date: December 19, 2014
Company: GE Healthcare
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact GE Healthcare directly.

Affected Products

GE Healthcare Automatic Mobile X-Ray (AMX) Series: Optima XR200 amx

Quantity: 6 units in US

Why Was This Recalled?

An issue was identified with the mAs accuracy check performed on the system that could result in inaccuracies in the 10% to 20% range for exposure times longer than 15 msec.

Where Was This Sold?

This product was distributed to 5 states: AZ, FL, GA, MO, WI

Affected (5 states)Not affected

About GE Healthcare

GE Healthcare has 88 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report