Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

GE Proteus XR/A radiographic system Recalled by GE Healthcare (China) Co., Ltd. Due to Steel cables in the Wall Stand for certain...

Date: October 24, 2024
Company: GE Healthcare (China) Co., Ltd.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact GE Healthcare (China) Co., Ltd. directly.

Affected Products

GE Proteus XR/A radiographic system

Quantity: 956 systems

Why Was This Recalled?

Steel cables in the Wall Stand for certain Proteus XR/a systems serviced by GE HealthCare have not been replaced at the intervals detailed in the Preventative Maintenance Action List.

Where Was This Sold?

This product was distributed nationwide across all 50 states.