Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

DeltaVision OMX SR Imaging System Model Number: 29115476 The DeltaVision Recalled by GE Healthcare Biosciences Due to Non-conformance with the DeltaVision OMX SR, the interlocks...

Date: September 10, 2018
Company: GE Healthcare Biosciences
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact GE Healthcare Biosciences directly.

Affected Products

DeltaVision OMX SR Imaging System Model Number: 29115476 The DeltaVision OMX SR Imaging System is a piece of laboratory equipment for super resolution imaging offluorescently-la beling biological samples. Techniques such as 3D Structured Illumination, Total Internal Reflection Fluorescence and localization microscopy may be used to examine a variety of samples over long periods of time.

Quantity: 39

Why Was This Recalled?

Non-conformance with the DeltaVision OMX SR, the interlocks are not wired in a redundant fashion to make the interlock system single-fault tolerant.

Where Was This Sold?

This product was distributed nationwide across all 50 states.