Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Future Diagnostics STAT-IntraOperative-Intact-PTH (STAT-IO-I-PTH) Immunoassay kit REF 4K-IPT-00. in vitro Recalled by Future Diagnostics Solutions B.V. Due to Due to an issue of high coefficient of...

Date: December 6, 2022
Company: Future Diagnostics Solutions B.V.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Future Diagnostics Solutions B.V. directly.

Affected Products

Future Diagnostics STAT-IntraOperative-Intact-PTH (STAT-IO-I-PTH) Immunoassay kit REF 4K-IPT-00. in vitro diagnostic.

Quantity: 342 kits

Why Was This Recalled?

Due to an issue of high coefficient of variability (CV) problem caused by the accuspheres inside the Immunoassay kits.

Where Was This Sold?

This product was distributed to 7 states: AZ, CA, CO, LA, OH, PA, TX

Affected (7 states)Not affected