Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Fujifilm Medical Endoscope for the duodenum and upper G.I. tract. ED-530XT Recalled by Fujifilm Medical Systems U.S.A., Inc. Due to The suction button on the ED-530XT endoscope may...

Date: June 30, 2015
Company: Fujifilm Medical Systems U.S.A., Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Fujifilm Medical Systems U.S.A., Inc. directly.

Affected Products

Fujifilm Medical Endoscope for the duodenum and upper G.I. tract. ED-530XT, ED-250XT5, ED-250XL5, ED-450XT5, and ED-450XL5 Endoscope Operation Manuals.

Quantity: 256 units

Why Was This Recalled?

The suction button on the ED-530XT endoscope may become lodged (stuck) in the active position following the use of x-ray contrast agents during ERCP procedures.

Where Was This Sold?

Distributed in the US and Canada.