Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

FUJIFILM FDR Visionary Suite with CH-200. Model # CH-200 Recalled by Fujifilm Medical Systems U.S.A., Inc. Due to A potential issue in the X-ray tube ceiling...

Date: June 15, 2017
Company: Fujifilm Medical Systems U.S.A., Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Fujifilm Medical Systems U.S.A., Inc. directly.

Affected Products

FUJIFILM FDR Visionary Suite with CH-200. Model # CH-200, Catalog: Visionary Suite (FMSU Marketing Name) The FDR Visionary Suite is intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations.

Quantity: 3

Why Was This Recalled?

A potential issue in the X-ray tube ceiling unit CH-200 which constitutes the Digital Radiography System FDR Visionary Suite. The X-ray tube assembly in these devices is mounted to a tube mounting flange, which is part of the tube holding shaft. where cracks may occur over time on the tube holding shaft near the base of the tube mounting flange.

Where Was This Sold?

US Distribution.

About Fujifilm Medical Systems U.S.A., Inc.

Fujifilm Medical Systems U.S.A., Inc. has 37 total recalls tracked by RecallDetector.