The device is a mobile x-ray system designed to work Recalled by FUJIFILM Healthcare Americas Corporation Due to When the key switch is turned on, the...

Date: November 23, 2023
Company: FUJIFILM Healthcare Americas Corporation
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact FUJIFILM Healthcare Americas Corporation directly.

Affected Products

The device is a mobile x-ray system designed to work with Fujifilm's GOS and CsI scintillator FDR D-EVO2 (DR-ID) 12XXSE) family of digital x-ray detectors coupled.

Quantity: 130

Why Was This Recalled?

When the key switch is turned on, the exposure conditions are set to the default values (85kV, 1.6 mAs, approx. 0.13mGy@ SID 1m) intended for adult chest imaging. In addition, if exposure conditions are set from the console while the key switch is off, the set exposure conditions will not be reflected.