Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

US AquaA 900H-Typ USA-3 x 208v-Intended for the purification of Recalled by Fresenius Medical Care Holdings, Inc. Due to Concentrate Flow was exceeding the maximum 19.9 bar...

Date: May 16, 2024
Company: Fresenius Medical Care Holdings, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Fresenius Medical Care Holdings, Inc. directly.

Affected Products

US AquaA 900H-Typ USA-3 x 208v-Intended for the purification of water to be used for hemodialysis Vivonic Part Numbers: G02020110 Fresenius Part Number: 24-090A-1

Quantity: 4 units

Why Was This Recalled?

Concentrate Flow was exceeding the maximum 19.9 bar set point. The standard default set point range is between 4 and 18.Error message (F020104, "concentrate pressure alarm limit exceeded") concentrate pressure is too high and stops the dialysis water supply.