Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Liberty Select Cycler (SW v.2.8.7) Recalled by Fresenius Medical Care Renal Therapies Group, LLC Due to The recalling firm identified a software issue related...

Date: January 24, 2018
Company: Fresenius Medical Care Renal Therapies Group, LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Fresenius Medical Care Renal Therapies Group, LLC directly.

Affected Products

Liberty Select Cycler (SW v.2.8.7), Material Number RTLR108343 Product Usage: The device is indicated for acute and chronic peritoneal dialysis.

Quantity: 9293

Why Was This Recalled?

The recalling firm identified a software issue related to the Patient Line Check (PLC) which may result in an increased risk of Overfill (also known as Increased Intraperitoneal Volume, IIPV). Overfill/IIPV may result in serious injury or death.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Fresenius Medical Care Renal Therapies Group, LLC

Fresenius Medical Care Renal Therapies Group, LLC has 27 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report