Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Liberty Select Cycler Software Version 2.9.0 Model Number 180343 and Recalled by Fresenius Medical Care Holdings, Inc. Due to The device may detect an incorrect Heater Bag...

Date: September 12, 2020
Company: Fresenius Medical Care Holdings, Inc.
Status: Completed
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Fresenius Medical Care Holdings, Inc. directly.

Affected Products

Liberty Select Cycler Software Version 2.9.0 Model Number 180343 and RTLR180343

Quantity: 9,907 devices

Why Was This Recalled?

The device may detect an incorrect Heater Bag volume which may lead to a 'Supply Bag Line Blocked' alarm during treatment. This alarm may cause the Heater Bag to contain more fluid than expected and the cycler does not recognize the extra fluid is present and it will not make it available for treatment. Not having enough solution to complete treatment may lead to inadequate dialysis which could result in the accumulation of toxins and fluid. Symptoms may include shortness of breath, increased weight gain, and increased blood pressure.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Fresenius Medical Care Holdings, Inc.

Fresenius Medical Care Holdings, Inc. has 161 total recalls tracked by RecallDetector.

Related Recalls