Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Honeywell Pressure Transducer- Part Number: 370506-22. It is a component Recalled by Fresenius Medical Care Holdings, Inc. Due to The firm is replacing pressure transducers with an...

Date: January 29, 2020
Company: Fresenius Medical Care Holdings, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Fresenius Medical Care Holdings, Inc. directly.

Affected Products

Honeywell Pressure Transducer- Part Number: 370506-22. It is a component used in 2008T, Common Name: Hemodialysis Delivery System; 2008T HD SYS. BLUE STAR; 2008K2, Common Name: Hemodialysis Delivery Equipment; 2008K@home, Common Name: Hemodialysis Delivery System

Quantity: 3561 units

Why Was This Recalled?

The firm is replacing pressure transducers with an improved design due to an increase in customer complaints regarding lower long-term reliability of the part.

Where Was This Sold?

This product was distributed nationwide across all 50 states.