Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Fresenius Liberty Cycler-Peritoneal Dialysis Cycler Model: 180111 and RTLR180111 Recalled by Fresenius Medical Care Renal Therapies Group, LLC Due to The door latch may not fully close and...

Date: December 15, 2015
Company: Fresenius Medical Care Renal Therapies Group, LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Fresenius Medical Care Renal Therapies Group, LLC directly.

Affected Products

Fresenius Liberty Cycler-Peritoneal Dialysis Cycler Model: 180111 and RTLR180111

Quantity: 14,094 units

Why Was This Recalled?

The door latch may not fully close and the door may open unexpectedly. The door may make physical contact with anything near the door of the machine, including the patient. May cause injury, delay in treatment, or Increased Intraperitoneal Volume (IIPV) (overfill).

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Fresenius Medical Care Renal Therapies Group, LLC

Fresenius Medical Care Renal Therapies Group, LLC has 27 total recalls tracked by RecallDetector.

Related Recalls