Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

AquaC UNO H [120V Recalled by Fresenius Medical Care Holdings, Inc. Due to Internal ethernet cable may be misaligned, the cable...

Date: June 16, 2022
Company: Fresenius Medical Care Holdings, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Fresenius Medical Care Holdings, Inc. directly.

Affected Products

AquaC UNO H [120V, 60Hz, 1000VA, Fresenius US-Version: 24-10VA-1 Vivonic Part Number: G02000970-US

Quantity: 318 units

Why Was This Recalled?

Internal ethernet cable may be misaligned, the cable jacket (insulation) could become compromised due to high temperature and result in the internal ethernet cable conducting electrical current to the ethernet port on the outside of the device), potentially harming service technicians and/or other authorized operators. If the ethernet port on the device is touched or manipulated, there is a risk of electrical shock

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Fresenius Medical Care Holdings, Inc.

Fresenius Medical Care Holdings, Inc. has 161 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report