Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

AquaBplus Water Purification System Recalled by Fresenius Medical Care Holdings, Inc. Due to AquaBplus HF module on the AquaBplus 2000 osmosis...

Date: August 3, 2021
Company: Fresenius Medical Care Holdings, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Fresenius Medical Care Holdings, Inc. directly.

Affected Products

AquaBplus Water Purification System, Part Number 24-2000-0, AquaBplus 2500 - Typ USA (G02040108-US), AquaBplus 2000 - Typ USA (G02040114-US), AquaBplus HF -Typ USA (G02040120FMC-US)

Quantity: 2 units

Why Was This Recalled?

AquaBplus HF module on the AquaBplus 2000 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2000 device. The 3-ring main and fluid fly loops the heat disinfection may not be as effective as a 2-ring main installation