Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

2008 Series Hemodialysis Machines: 2008Kathome - 190895 2008T GEN 2 Recalled by Fresenius Medical Care Holdings, Inc. Due to The Acetate value for GranuFlo on the Select...

Date: February 20, 2015
Company: Fresenius Medical Care Holdings, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Fresenius Medical Care Holdings, Inc. directly.

Affected Products

2008 Series Hemodialysis Machines: 2008Kathome - 190895 2008T GEN 2 BIBAG WITHOUT CDX; 190908 2008T HD SYSTEM WITH CDX, CANADA; 190713 2008T HEMODIALYSIS SYS., with CDX; 190766 2008T HEMODIALYSIS SYSTEM W/BIBAG; 190858 2008T HEMODIALYSIS SYSTEM WITHOUT CDX The Fresenius 2008K is indicated for acute and chronic dialysis therapy.

Quantity: 133,751 in total (US: 130,337 and OUS: Canada 1,201 and Mexico 2,213)

Why Was This Recalled?

The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen.