Ivenix Infusion System (IIS) Recalled by Fresenius Kabi USA, LLC Due to Large Volume Pump Software, version 5.9.2 and earlier...

Date: January 10, 2025
Company: Fresenius Kabi USA, LLC
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Fresenius Kabi USA, LLC directly.

Affected Products

Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration.

Quantity: 23 systems

Why Was This Recalled?

Large Volume Pump Software, version 5.9.2 and earlier has potential for the following anomalies: 1. The pump may become nonfunctional if during an alarm condition the Pause Audio option is repeated 70 time or more. 2. If a secondary infusion is started at the exact moment a primary infusion completes and VTBI reaches 0, it will switch to primary. The primary infusion will infuse at the previously programed primary rate and continue until the infusion is stopped or the bag is empty.

Where Was This Sold?

This product was distributed to 15 states: CA, CO, GA, ID, MI, MS, NE, NV, NJ, OK, TX, UT, VA, WA, WI

Affected (15 states)Not affected

About Fresenius Kabi USA, LLC

Fresenius Kabi USA, LLC has 67 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report