Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

PT301US Airvo 3 Respiratory Support Device with software version 1.2.0 Recalled by Fisher & Paykel Healthcare, Ltd. Due to Due to a software issue, affected devices that...

Date: August 7, 2024
Company: Fisher & Paykel Healthcare, Ltd.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Fisher & Paykel Healthcare, Ltd. directly.

Affected Products

PT301US Airvo 3 Respiratory Support Device with software version 1.2.0 and/or 1.5.1 The Airvo 3 is intended to provide high flow warmed and humidified respiratory gases for administration to spontaneously breathing infant, child, adolescent and adult patients in hospitals and sub-acute facilities.

Quantity: 944 units

Why Was This Recalled?

Due to a software issue, affected devices that are set up with High Pressure Oxygen (HPO), if the flow alignment alarm occurs, the device will deliver room air only. If this happens, a patient my experience oxygen desaturation that could lead to hypoxia.

Where Was This Sold?

This product was distributed to 29 states: AZ, CA, CO, CT, FL, GA, IL, IA, KY, MD, MA, MI, MN, MO, MT, NV, NH, NJ, NY, ND, OH, OR, PA, SD, TX, UT, VA, WA, DC

Affected (29 states)Not affected

About Fisher & Paykel Healthcare, Ltd.

Fisher & Paykel Healthcare, Ltd. has 7 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report