Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Logic Fit Tibial Tamp Head Recalled by Exactech, Inc. Due to The firm determined that the labeling (surgical technique)...

Date: May 25, 2017
Company: Exactech, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Exactech, Inc. directly.

Affected Products

Logic Fit Tibial Tamp Head

Quantity: 573

Why Was This Recalled?

The firm determined that the labeling (surgical technique) should be updated to clarify the proper technique and instrumentation to remove the tamp assembly from the bone, i.e., to include a caution statement about the potential for instrument breakage if the tamp handle/guide is misused by impacting in retro-grade.

Where Was This Sold?

This product was distributed to 30 states: AL, AZ, CA, CO, FL, GA, ID, IL, IN, IA, KS, LA, ME, MD, MA, MN, NV, NJ, NY, NC, OH, OK, OR, RI, SC, TN, TX, VA, WA, WI