Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Exact Sciences Sample Mixer 2. For In Vitro Diagnostic Use. Recalled by Exact Sciences Corporation Due to Exact Sciences is correcting all Sample Mixer v2...

Date: April 28, 2017
Company: Exact Sciences Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Exact Sciences Corporation directly.

Affected Products

Exact Sciences Sample Mixer 2. For In Vitro Diagnostic Use. The Exact Sciences Sample Mixer 2 is intended for mixing stool samples in a Sample Container as part of pre-analytic sample processing for in vitro diagnostic tests

Quantity: 15

Why Was This Recalled?

Exact Sciences is correcting all Sample Mixer v2 Upgrade units because of incidents of Sample Mixer door damage, which has occurred when the Sample Holder opens during a run and the Sample Holder lid makes contact with the Sample Mixer door.

Where Was This Sold?

This product was distributed to 1 state: WI

Affected (1 state)Not affected

About Exact Sciences Corporation

Exact Sciences Corporation has 1 total recall tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report