Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

SlideView VS200 slide viewer Recalled by EVIDENT SCIENTIFIC INC Due to A broken fuse inside the laser that was...

Date: July 22, 2024
Company: EVIDENT SCIENTIFIC INC
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact EVIDENT SCIENTIFIC INC directly.

Affected Products

SlideView VS200 slide viewer,

Quantity: N/A

Why Was This Recalled?

A broken fuse inside the laser that was caused by a short circuit at PIN 4 of the DOOR CTRL. This short circuit was caused by a badly positioned cable that made contact with the housing/frame and caused a short circuit and destroyed the fuse in the laser. This caused the shutter to remain in its most recent state; either in an open or closed position. Once the fuse was destroyed the shutter could not be operated.