Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
MONOCRYL (poliglecaprone 25) Suture - Intended for use in general Recalled by Ethicon, Inc. Due to Issue on a specific packaging machine resulted in...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Ethicon, Inc. directly.
Affected Products
MONOCRYL (poliglecaprone 25) Suture - Intended for use in general soft tissue approximation and/or ligation where an absorbable material is indicated PRODUCT CODE: Y315H Y345H Y359H Y398H Y416H Y417H Y426H Y427H Y935H Y936H Y945H
Quantity: 306000 units
Why Was This Recalled?
Issue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Ethicon, Inc.
Ethicon, Inc. has 69 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report