Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Ligamax 5 mm Endoscopic Multiple Clip Applier Recalled by Ethicon Endo-Surgery Inc Due to A potential exists that certain LIGAMAX devices within...

Date: November 7, 2019
Company: Ethicon Endo-Surgery Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Ethicon Endo-Surgery Inc directly.

Affected Products

Ligamax 5 mm Endoscopic Multiple Clip Applier

Quantity: 48,428

Why Was This Recalled?

A potential exists that certain LIGAMAX devices within the impacted lots may have small holes in the Tyvek lidding which could result in a breach of sterility.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Ethicon Endo-Surgery Inc

Ethicon Endo-Surgery Inc has 54 total recalls tracked by RecallDetector.