Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

AC/DC POWER SUPPLY for 7400 Recalled by ESAOTE S.P.A Due to The AC/DC adapter (power supply) of the portable...

Date: September 8, 2020
Company: ESAOTE S.P.A
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact ESAOTE S.P.A directly.

Affected Products

AC/DC POWER SUPPLY for 7400, Part Number 151004300, XP Power Model Number AHM180PS19-XE0338, AC/DC Adapter associated with the following portable ultrasound diagnostic systems: The AC/DC adapter is intended for use with the following portable ultrasound systems: MyLabAlpha (101740000) MyLabGamma (101741000) MyLabOmega (101740004) MyLabSigma (101741004) MyLabSigma elite (101741007) MyLabAlphaVET (101740001) MyLabGammaVET (101741001) MyLabDelta (100743000) MyLabDeltaVET (100743001) MyLabOmegaVET (101740005) MyLabSigmaVET (101741005)

Quantity: 989 total adapters in the US

Why Was This Recalled?

The AC/DC adapter (power supply) of the portable ultrasound systems listed below that may result in the connector heating up, generating smoke and melting. This condition is caused by the mechanical stress of the connection between the AC/DC power adapter and the mains cable over the time of use.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About ESAOTE S.P.A

ESAOTE S.P.A has 2 total recalls tracked by RecallDetector.