Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
AC/DC POWER SUPPLY for 7400 Recalled by ESAOTE S.P.A Due to The AC/DC adapter (power supply) of the portable...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact ESAOTE S.P.A directly.
Affected Products
AC/DC POWER SUPPLY for 7400, Part Number 151004301, XP Power Model Number AHM180PS19-XE1057, AC/DC Adapter associated with the following portable ultrasound diagnostic systems: The AC/DC adapter is intended for use with the following portable ultrasound systems: MyLabAlpha (101740000) MyLabGamma (101741000) MyLabOmega (101740004) MyLabSigma (101741004) MyLabSigma elite (101741007) MyLabAlphaVET (101740001) MyLabGammaVET (101741001) MyLabDelta (100743000) MyLabDeltaVET (100743001) MyLabOmegaVET (101740005) MyLabSigmaVET (101741005)
Quantity: 989 total adapters in the US
Why Was This Recalled?
The AC/DC adapter (power supply) of the portable ultrasound systems listed below that may result in the connector heating up, generating smoke and melting. This condition is caused by the mechanical stress of the connection between the AC/DC power adapter and the mains cable over the time of use.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About ESAOTE S.P.A
ESAOTE S.P.A has 2 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report