Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

epoc Blood Analysis BGEM Test Cards for use with the Recalled by Epocal Inc Due to Customer reported problem while performing incoming QC of...

Date: June 28, 2011
Company: Epocal Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Epocal Inc directly.

Affected Products

epoc Blood Analysis BGEM Test Cards for use with the epoc(R) Blood Analysis System Product Usage: Usage: epoc Blood Analysis BGEM Test Cards are used with the epoc(R) Blood Analysis System which is a POC testing system

Quantity: 8,650

Why Was This Recalled?

Customer reported problem while performing incoming QC of newly received lot of test cards. Liquid quality control results were found to be out of range. Further investigation by Epocal determined that a portion of the test card lot (approximately 3000 cards) could potentially report low glucose results.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected