Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

epoc BGEM Test Card (Catalogue number CT-1004-00-00). quantitative testing of Recalled by Epocal Due to This recall is being initiated because the affected...

Date: March 13, 2013
Company: Epocal
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Epocal directly.

Affected Products

epoc BGEM Test Card (Catalogue number CT-1004-00-00). quantitative testing of samples for pH, pCO2, pO2, Sodium Potassium , Ionized Calcium, Glucose, Lactate and Hematocrit.

Quantity: 828,750 Test Card

Why Was This Recalled?

This recall is being initiated because the affected lots of this product may yield results that are not consistent with stated precision performance criteria at concentration values at the high-end of the Glucose test sensor (>400 mg/dl). Specifically, after a period of storage of three (3) to four (4) months, the Glucose sensor may report low results at the high end of the measurement range.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected