Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

REF 800-05-035 Recalled by Enztec Limited Due to The femoral impactors have a higher than normal...

Date: August 16, 2019
Company: Enztec Limited
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Enztec Limited directly.

Affected Products

REF 800-05-035, ENZTEC FEMORAL IMPACTOR - DJO, (01) 09421028110624 (10) 999999, To be used only with DJO EMPOWR 3D KNEE

Quantity: 312 Femoral Impactors

Why Was This Recalled?

The femoral impactors have a higher than normal risk of the arm breaking. During impaction the tip of the arm may fracture and separate from the instrument. This may cause a minor delay in surgery. If not retrieved, it may cause an inflammatory response that may require revision surgery.

Where Was This Sold?

This product was distributed to 1 state: CA

Affected (1 state)Not affected

About Enztec Limited

Enztec Limited has 1 total recall tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report