Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Turon Shoulder Impaction Fixture REF 804-15-102 The Turon and RSP Recalled by Encore Medical, Lp Due to During the Turon assembly, the impaction forces caused...

Date: September 14, 2015
Company: Encore Medical, Lp
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Encore Medical, Lp directly.

Affected Products

Turon Shoulder Impaction Fixture REF 804-15-102 The Turon and RSP Impaction Fixtures are used in the operating room on the back table to assemble the shoulder systems for implant. The Turon Fixture is used to connect the Humeral Head to the Humeral Neck, then subsequently to assemble the Humeral Stem to the Head/Neck construct. The RSP Fixture is used to assemble the UHMWPE Socket Inserts to both the RSP Socket Shells and Monoblock Stems.

Quantity: 277 units

Why Was This Recalled?

During the Turon assembly, the impaction forces caused the polymer, black acetal copolymer form the Impaction Fixture to wear off on the lateral surface of the Humeral Stem titanium plasma spray coating.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Encore Medical, Lp

Encore Medical, Lp has 43 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report