Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
MODIFIED/SPECIAL INSTRUMENT Recalled by Encore Medical, Lp Due to It was reported that the models used to...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Encore Medical, Lp directly.
Affected Products
MODIFIED/SPECIAL INSTRUMENT, TURON GLENOID REAMER, REAM AND RUN, Catalog Numbers: REF S-200956 - 5mm PIN, SIZE 46 REF S-200957 - 5mm PIN, SIZE 50 REF S-200958 - 5mm PIN, SIZE 58
Quantity: 5 units
Why Was This Recalled?
It was reported that the models used to manufacture S-200956/958 had the wrong dimension used to create the spherical radius cutting surface. Device design allows a larger spherical radius cutting surface than intended.
Where Was This Sold?
This product was distributed to 2 states: OH, VA
About Encore Medical, Lp
Encore Medical, Lp has 43 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report