Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

MODIFIED/SPECIAL INSTRUMENT Recalled by Encore Medical, Lp Due to It was reported that the models used to...

Date: August 4, 2017
Company: Encore Medical, Lp
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Encore Medical, Lp directly.

Affected Products

MODIFIED/SPECIAL INSTRUMENT, TURON GLENOID REAMER, REAM AND RUN, Catalog Numbers: REF S-200956 - 5mm PIN, SIZE 46 REF S-200957 - 5mm PIN, SIZE 50 REF S-200958 - 5mm PIN, SIZE 58

Quantity: 5 units

Why Was This Recalled?

It was reported that the models used to manufacture S-200956/958 had the wrong dimension used to create the spherical radius cutting surface. Device design allows a larger spherical radius cutting surface than intended.

Where Was This Sold?

This product was distributed to 2 states: OH, VA

Affected (2 states)Not affected

About Encore Medical, Lp

Encore Medical, Lp has 43 total recalls tracked by RecallDetector.

Related Recalls