Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
EMPOWR VVC KNEE TIBIAL INSERT Recalled by Encore Medical, LP Due to Tibial insert implant large boxes were labeled small...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Encore Medical, LP directly.
Affected Products
EMPOWR VVC KNEE TIBIAL INSERT, SIZE 6, 22MM, REF: 346-22-706; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 7, 22MM, REF: 346-22-707; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 8, 22MM, REF: 346-22-708; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 9, 22MM, REF: 346-22-709; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 10, 22MM, REF: 346-22-710; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 11, 22MM, REF: 346-22-711; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 5 BRIDGE UP, 22MM, REF: 346-22-755; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 6 BRIDGE DOWN, 22MM, REF: 346-22-766; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 2, 25MM, REF: 346-25-702; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 3, 25MM, REF: 346-25-703; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 4, 25MM, REF: 346-25-704; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 5, 25MM, REF: 346-25-705; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 5 BRIDGE UP, 25MM, REF: 346-25-755; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 6 BRIDGE DOWN, 25MM, REF: 346-25-766
Quantity: 216
Why Was This Recalled?
Tibial insert implant large boxes were labeled small and the small boxes were labeled large. If the incorrect labeling is identified this could lead to surgical delay and if the surgeon does not recognize the size discrepancy between the insert label and the pin label, the surgeon may implant the incorrect pin size.
Where Was This Sold?
This product was distributed to 5 states: FL, IL, IN, MN, NY
About Encore Medical, LP
Encore Medical, LP has 54 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report