Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Altivate Anatomic(TM) INSTRUMENTATION Recalled by Encore Medical, Lp Due to It was discovered during a surgery that the...

Date: January 29, 2018
Company: Encore Medical, Lp
Status: Completed
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Encore Medical, Lp directly.

Affected Products

Altivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 50mm, REF 804-07-500 Product Usage: The AltiVate Anatomic Shoulder System is a total shoulder system indicated as an anatomic shoulder joint replacement for patients suffering from pain and dysfunction.

Quantity: 92 units

Why Was This Recalled?

It was discovered during a surgery that the design of the radius at the base of the pegs of the Anatomic Glenoid Trial (Part Numbers: 804-07-380, 804-07-420, 804-07-460, 804-07-500, and 804-07-540) is much larger on the trails than on the implants

Where Was This Sold?

This product was distributed nationwide across all 50 states.