XiO RPT System The XiO RTP System is used to Recalled by Elekta, Inc. Due to Dose and Monitor Unit values are not computed...

Date: February 17, 2014
Company: Elekta, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Elekta, Inc. directly.

Affected Products

XiO RPT System The XiO RTP System is used to create treatment plans for any cancer patient for whom external beam radiation therapy or brachytherapy has been prescribed. The system will calculate and display, both on screen and in hard-copy, either two or three dimensional radiation dose distributions within a patient for a given treatment.

Quantity: 1776

Why Was This Recalled?

Dose and Monitor Unit values are not computed correctly when Elekta Motorized Wedges are used with Dynamic Conformal arcs.

Where Was This Sold?

This product was distributed to 41 states: AL, AK, AZ, CA, CO, CT, DE, FL, GA, IL, IN, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NJ, NY, NC, ND, OH, OK, PA, RI, SC, TN, TX, VA, WA, WV, WI, WY, DC