Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Precise Digital Accelerator Delivery of radiation to defined target volumes Recalled by Elekta, Inc. Due to There have been a small number of instances...

Date: August 6, 2012
Company: Elekta, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Elekta, Inc. directly.

Affected Products

Precise Digital Accelerator Delivery of radiation to defined target volumes

Quantity: 110

Why Was This Recalled?

There have been a small number of instances in hospitals where the joint between the drive gear and the gantry base drive wheel has become loose, leading to fatigue failure of some of the securing bolts.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Elekta, Inc.

Elekta, Inc. has 100 total recalls tracked by RecallDetector.

Related Recalls