Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Oncentra External Beam - VMAT - Radiation therapy planning system Recalled by Elekta, Inc. Due to When using the option "Tumor Overlap Fraction" in...

Date: December 18, 2015
Company: Elekta, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Elekta, Inc. directly.

Affected Products

Oncentra External Beam - VMAT - Radiation therapy planning system Product Usage: The Oncentra system is a radiation treatment planning software designed to analyze and plan radiation treatments in three dimensions for the purpose of treating patients with cancer.

Quantity: 154

Why Was This Recalled?

When using the option "Tumor Overlap Fraction" in VMAT planning it has been observed that in rare cases the system does include an organ at risk as target volume. This could result in open MLC, and open jaws in areas away from the target volume.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Elekta, Inc.

Elekta, Inc. has 100 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report