Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

MOSAIQ Oncology Information System. Used to manage workflows for treatment Recalled by Elekta, Inc. Due to First Databank (FDB) is a third party product...

Date: November 24, 2015
Company: Elekta, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Elekta, Inc. directly.

Affected Products

MOSAIQ Oncology Information System. Used to manage workflows for treatment planning and delivery.

Quantity: 211

Why Was This Recalled?

First Databank (FDB) is a third party product incorporated into MOSAIQ. FDB routinely modifies drugs in their database to reflect market and formulary changes. In FDB Framework v2.0, those drugs are not being included in drug-drug, drug-allergy and duplicate drug therapy interaction checking.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Elekta, Inc.

Elekta, Inc. has 100 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report